ENDOTRACHEAL TUBE CUFF PRESSURE CONTROL DURING ANESTHESIA IN CANCER PATIENTS

N.R. Abdukhalilov 1, A.A. Arynov 1, D.A. Baidaulet 1, A.A.Nurmanova 1, E.A. Seidalieva 1, V.V. Chursin 2

1. «Kazakh Research Institute of Oncology and Radiology» JSC, Аlmaty, the Republic of Kazakhstan;
2. Kazakhstan Medical University «Higher School of Public Healthcare», Аlmaty, the Republic of Kazakhstan

DOI: https://www.doi.org/10.52532/2521-6414-2022-4-66-38-41

UDC: 616-006-089-07

Year: 2022 issure: 66 number: 4 pages: 38-41

Download PDF: 2012.2-3.24-25_10.pdf

ABSTRACT

Relevance: An endotracheal tube (ETT) cuff provides a seal and encloses the lower airway from aspiration. Normally, the pressure in the ETT cuff is in the range of 20 to 30 cm of water column. Both increased and insufficient inflation of the ETT cuff is associated with a number of complications.
The purpose of the research: comparison of palpation and apparatus methods of pressure control in the cuff of the endotracheal tube during anesthesia in cancer patients.
Methods: A prospective observational study included 60 patients during general anesthesia in the department of anesthesiology and intensive care of KazIOR. Air was injected into the ETT cuff using a syringe, followed by palpation of the ETT cuff balloon and pressure control using the IntelliCuff device (Hamilton Medical, Switzerland). The actual pressure was compared with normal values, then the volume of air, actual and necessary to achieve normal pressure, was estimated.
Results: assessment of the pressure level by the “classic” palpation method led to errors in the pressure level in the ETT cuff in more than 50% of cases; the normal level of pressure was only in 25 patients (42%), while the measured air volume in the cuff was on average 5.9±1.9 ml, although for the average air volume to achieve a pressure of 25 mm of water column was 3.9 ml, which led to an overestimated level of pressure in the ETT cuff.
Conclusion: Determining ETT cuff pressure by palpation of the control balloon, while common practice, often results in incorrect pressure readings. At the same time, both high and low pressure in the ETT cuff are associated with the development of complications. The use of devices for measuring pressure in the ETT cuff allows you to control its level, while devices that allow prolonged monitoring of pressure in the ETT cuff have an advantage.
Keywords: endotracheal tube, cuff pressure control, microaspiration, tracheal intubation, lung ventilation.

List of sources used:

  1. Szmuk P., Ezri T., Evron S., Roth Y., Katz J. A brief history of tracheostomy and tracheal intubation, from the Bronze Age to the Space Age // Intensive Care Med. – 2008. – Vol. 34(2). – P. 222-228. https://doi.org/10.1007/s00134-007-0931-5
  2. Khan M.U., Khokar R., Qureshi S., Al Zahrani T., Aqil M., Shiraz M. Measurement of endotracheal tube cuff pressure: Instrumental versus conventional method // Saudi J. Anaesth. – 2016. – Vol. 10 (4). – P. 428-431. https://doi.org/10.4103/1658-354X.179113
  3. Harm F., Zuercher M., Bassi M., Ummenhofer W. Prospective observational study on tracheal tube cuff pressures in emergency patients – is neglecting the problem the problem? // Scand. J. Trauma Resusc. Emerg. Med. – 2013. – Vol. 21. – P. 83. https://doi.org/10.1186/1757-7241-21-83
  4. Seyed Siamdoust S.A., Mohseni M., Memarian A. Endotracheal Tube Cuff Pressure Assessment: Education May Improve but not Guarantee the Safety of Palpation Technique // Anesth. Pain Med. – 2015. – Vol. 5(3). – Art. no. e16163. https://doi.org/10.5812/aapm.5(3)2015.16163
  5. Rosero E.B., Ozayar E., Eslava-Schmalbach J., Minhajuddin A., Joshi G.P. Effects of Increasing Airway Pressures on the Pressure of the Endotracheal Tube Cuff During Pelvic Laparoscopic Surgery // Anesth. Analg. – 2018. – Vol. 127(1). – P. 120-125. https://doi.org/10.1213/ANE.0000000000002657
  6. Muniappan A., Wain J.C., Wright C.D., Donahue D.M., Gaissert H., Lanuti M., Mathisen D.J. Surgical treatment of nonmalignant tracheoesophageal fistula: a thirty-five-year experience // Ann. Thorac. Surg. – 2013. – Vol. 95(4). – P. 1141-1146. https://doi.org/10.1016/j.athoracsur.2012.07.041
  7. Ganason N., Sivanaser V., Liu C.Y., Maaya M., Ooi J.S.M. Post-operative Sore Throat: Comparing the Monitored Endotracheal Tube Cuff Pressure and Pilot Balloon Palpation Methods // Malays. J. Med. Sci. – 2019. – Vol. 26(5). – P. 132-138. https://doi.org/10.21315/mjms2019.26.5.12
  8. Maertens B., Lin F., Chen Y., Rello J., Lathyris D., Blot S. Effectiveness of Continuous Cuff Pressure Control in Preventing Ventilator-Associated Pneumonia: A Systematic Review and Meta-Analysis of Randomized Controlled Trials // Crit. Care Med. – 2022. – Vol. 50(10). – P. 1430-1439. https://doi.org/10.1097/CCM.0000000000005630
  9. Blot S., Ruppé E., Harbarth S., Asehnoune K., Poulakou G., Luyt C.E., Rello J., Klompas M., Depuydt P., Eckmann C., Martin-Loeches I., Povoa P., Bouadma L., Timsit J.F., Zahar J.R. Healthcare-associated infections in adult intensive care unit patients: Changes in epidemiology, diagnosis, prevention and contributions of new technologies // Intensive Crit. Care Nurs. – 2022. – Vol. 70. – Art. no. 103227. https://doi.org/10.1016/j.iccn.2022.103227
  10. Modi A.R., Kovacs C.S. Hospital-acquired and ventilator-associated pneumonia: Diagnosis, management, and prevention // Cleve Clin. J. Med. – 2020. – Vol. 87(10). – P. 633-639. https://doi.org/10.3949/ccjm.87a.19117
  11. Lizy C., Swinnen W., Labeau S., Poelaert J., Vogelaers D., Vandewoude K., Dulhunty J., Blot S. Cuff pressure of endotracheal tubes after changes in body position in critically ill patients treated with mechanical ventilation // Am. J. Crit. Care. – 2014. – Vol. 23(1). – P. e1-8. https://doi.org/10.4037/ajcc2014489
  12. Danielis M., Benatti S., Celotti P., De Monte A., Trombini O. Il monitoraggio pressorio continuo della cuffia del tubo endotracheale: best practice in terapia intensive [Continuous monitoring of endotracheal tube cuff pressure: best practice in intensive care unit (in Italian)] // Assist. Inferm. Ric. – 2015. – Vol. 34(1). – P. 15-20. https://doi.org/10.1702/1812.19746
  13. Dauvergne J.E., Geffray A.L., Asehnoune K., Rozec B., Lakhal K. Automatic regulation of the endotracheal tube cuff pressure with a portable elastomeric device. A randomised controlled study // Anaesth. Crit. Care Pain Med. – 2020. – Vol. 39(3). – P. 435-441. https://doi.org/10.1016/j.accpm.2020.04.007
  14. Detsky M.E., Jivraj N., Adhikari N.K., Friedrich J.O., Pinto R., Simel D.L., Wijeysundera D.N., Scales D.C. Will This Patient Be Difficult to Intubate?: The Rational Clinical Examination Systematic Review // JAMA. – 2019. – Vol. 321(5). – P. 493-503. https://doi.org/10.1001/jama.2018.21413
  15. Kumar C.M., Seet E., Van Zundert T.C.R.V. Measuring endotracheal tube intracuff pressure: no room for complacency // J. Clin. Monit. Comput. – 2021. – Vol. 35(1). – P. 3-10. https://doi.org/10.1007/s10877-020-00501-2
  16. Coelho R. de M., de Paiva T.T., da Silva Telles Mathias L.A. In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes // Braz. J. Anesthesiol. – 2016. – Vol. 66(2). – P. 120-125. https://doi.org/10.1016/j.bjane.2014.06.012
  17. Duarte N.M.D.C, Caetano A.M.M., Arouca G.O., Ferrreira A.T., Figueiredo J.L. Insuflação de balonete de tubo traqueal por método subjetivo: desempenho de médicos residentes e especialistas em anestesiologia. Estudo prospectivo observacional [Subjective method for tracheal tube cuff inflation: performance of anesthesiology residents and staff anesthesiologists. Prospective observational study (in Spanish)] // Braz. J. Anesthesiol. – 2020. – Vol. 70(1). – P. 9-14. https://doi.org/10.1016/j.bjan.2019.09.010
  18. Li M., Yiu Y., Merrill T., Yildiz V., deSilva B., Matrka L. Risk factors for posttracheostomy tracheal stenosis // Otolaryngol. Head Neck Surg. –2018. – Vol. 159(4). – P. 698-704. https://doi.org/10.1177/0194599818794456
  19. Abubaker .J, Zia Ullah S., Ahmed S., Rehman Memon A.U., Abubaker Z.J., Ansari M.I., Karim M. Evaluating the Knowledge of Endotracheal Cuff Pressure Monitoring Among Critical Care Providers by Palpation of Pilot Balloon and By Endotracheal Tube Cuff Manometer // Cureus. – 2019. – Vol. 11(7). – Art. no. e5061. https://doi.org/10.7759/cureus.5061
  20. Marjanovic N., Boisson M., Asehnoune K., Foucrier A., Lasocki S., Ichai C., Leone M., Pottecher J., Lefrant J.Y., Falcon D., Veber B., Chabanne R., Drevet C.M., Pili-Floury S., Dahyot-Fizelier C., Kerforne T., Seguin S., de Keizer J., Frasca D., Guenezan J., Mimoz O.; AGATE Study Group. Continuous Pneumatic Regulation of Tracheal Cuff Pressure to Decrease Ventilator-associated Pneumonia in Trauma Patients Who Were Mechanically Ventilated: The AGATE Multicenter Randomized Controlled Study // Chest. – 2021. – Vol. 160(2). – P. 499-508. https://doi.org/10.1016/j.chest.2021.03.007
  21. Sevdi M.S., Demirgan S., Erkalp K., Akyol O., Ozcan F.G., Guneyli H.C., Tunali M.C., Selcan A. Continuous Endotracheal Tube Cuff Pressure Control Decreases Incidence of Ventilator-Associated Pneumonia in Patients with Traumatic Brain Injury // J. Invest. Surg. – 2022. – Vol. 35(3). – P. 525-530. https://doi.org/10.1080/08941939.2021.1881190
  22. Chenelle C.T., Oto J., Sulemanji D., Fisher D.F., Kacmarek R.M. Evaluation of an automated endotracheal tube cuff controller during simulated mechanical ventilation // Respir. Care. – 2015. – Vol. 60(2). – P. 183-190. https://doi.org/10.4187/respcare.03387

Error: Contact form not found.