ENDOTRACHEAL TUBE CUFF PRESSURE CONTROL DURING ANESTHESIA IN CANCER PATIENTS
N.R. Abdukhalilov 1, A.A. Arynov 1, D.A. Baidaulet 1, A.A.Nurmanova 1, E.A. Seidalieva 1, V.V. Chursin 2
1. «Kazakh Research Institute of Oncology and Radiology» JSC, Аlmaty, the Republic of Kazakhstan;
2. Kazakhstan Medical University «Higher School of Public Healthcare», Аlmaty, the Republic of Kazakhstan
DOI: https://www.doi.org/10.52532/2663-4864-2022-4-66-38-41
UDC: 616-006-089-07
Year: 2022 issure: 66 number: 4 pages: 38-41
ABSTRACT
Relevance: An endotracheal tube (ETT) cuff provides a seal and encloses the lower airway from aspiration. Normally, the pressure in the ETT cuff is in the range of 20 to 30 cm of water column. Both increased and insufficient inflation of the ETT cuff is associated with a number of complications.
The study aimed to compare the palpation and apparatus methods of pressure control in the endotracheal tube cuff during anesthesia in cancer patients.
Methods: A prospective observational study included 60 patients during general anesthesia in the department of anesthesiology and intensive care of KazIOR. Air was injected into the ETT cuff using a syringe, followed by palpating the ETT cuff balloon and pressure control using the IntelliCuff device (Hamilton Medical, Switzerland). The actual pressure was compared with normal values, then the air volume, actual and necessary to achieve normal pressure, was estimated.
Results: Assessment of the pressure level by the “classic” palpation method led to errors in the pressure level in the ETT cuff in more than 50% of cases; the normal level of pressure was only in 25 patients (42%), while the measured air volume in the cuff was on average 5.9±1.9 mL, although for the average air volume to achieve a pressure of 25 mm of water column was 3.9 mL, which led to an overestimated level of pressure in the ETT cuff.
Conclusion: Determining ETT cuff pressure by palpating the control balloon is a common practice that often results in incorrect pressure readings. At the same time, both high and low pressure in the ETT cuff is associated with complications. Using devices for measuring pressure in the ETT cuff allows you to control its level, while devices that allow prolonged monitoring of pressure in the ETT cuff have an advantage.
Keywords: endotracheal tube, cuff pressure control, microaspiration, tracheal intubation, lung ventilation.
List of sources used:
- Szmuk P., Ezri T., Evron S., Roth Y., Katz J. A brief history of tracheostomy and tracheal intubation, from the Bronze Age to the Space Age // Intensive Care Med. – 2008. – Vol. 34(2). – P. 222-228. https://doi.org/10.1007/s00134-007-0931-5
- Khan M.U., Khokar R., Qureshi S., Al Zahrani T., Aqil M., Shiraz M. Measurement of endotracheal tube cuff pressure: Instrumental versus conventional method // Saudi J. Anaesth. – 2016. – Vol. 10 (4). – P. 428-431. https://doi.org/10.4103/1658-354X.179113
- Harm F., Zuercher M., Bassi M., Ummenhofer W. Prospective observational study on tracheal tube cuff pressures in emergency patients – is neglecting the problem the problem? // Scand. J. Trauma Resusc. Emerg. Med. – 2013. – Vol. 21. – P. 83. https://doi.org/10.1186/1757-7241-21-83
- Seyed Siamdoust S.A., Mohseni M., Memarian A. Endotracheal Tube Cuff Pressure Assessment: Education May Improve but not Guarantee the Safety of Palpation Technique // Anesth. Pain Med. – 2015. – Vol. 5(3). – Art. no. e16163. https://doi.org/10.5812/aapm.5(3)2015.16163
- Rosero E.B., Ozayar E., Eslava-Schmalbach J., Minhajuddin A., Joshi G.P. Effects of Increasing Airway Pressures on the Pressure of the Endotracheal Tube Cuff During Pelvic Laparoscopic Surgery // Anesth. Analg. – 2018. – Vol. 127(1). – P. 120-125. https://doi.org/10.1213/ANE.0000000000002657
- Muniappan A., Wain J.C., Wright C.D., Donahue D.M., Gaissert H., Lanuti M., Mathisen D.J. Surgical treatment of nonmalignant tracheoesophageal fistula: a thirty-five-year experience // Ann. Thorac. Surg. – 2013. – Vol. 95(4). – P. 1141-1146. https://doi.org/10.1016/j.athoracsur.2012.07.041
- Ganason N., Sivanaser V., Liu C.Y., Maaya M., Ooi J.S.M. Post-operative Sore Throat: Comparing the Monitored Endotracheal Tube Cuff Pressure and Pilot Balloon Palpation Methods // Malays. J. Med. Sci. – 2019. – Vol. 26(5). – P. 132-138. https://doi.org/10.21315/mjms2019.26.5.12
- Maertens B., Lin F., Chen Y., Rello J., Lathyris D., Blot S. Effectiveness of Continuous Cuff Pressure Control in Preventing Ventilator-Associated Pneumonia: A Systematic Review and Meta-Analysis of Randomized Controlled Trials // Crit. Care Med. – 2022. – Vol. 50(10). – P. 1430-1439. https://doi.org/10.1097/CCM.0000000000005630
- Blot S., Ruppé E., Harbarth S., Asehnoune K., Poulakou G., Luyt C.E., Rello J., Klompas M., Depuydt P., Eckmann C., Martin-Loeches I., Povoa P., Bouadma L., Timsit J.F., Zahar J.R. Healthcare-associated infections in adult intensive care unit patients: Changes in epidemiology, diagnosis, prevention and contributions of new technologies // Intensive Crit. Care Nurs. – 2022. – Vol. 70. – Art. no. 103227. https://doi.org/10.1016/j.iccn.2022.103227
- Modi A.R., Kovacs C.S. Hospital-acquired and ventilator-associated pneumonia: Diagnosis, management, and prevention // Cleve Clin. J. Med. – 2020. – Vol. 87(10). – P. 633-639. https://doi.org/10.3949/ccjm.87a.19117
- Lizy C., Swinnen W., Labeau S., Poelaert J., Vogelaers D., Vandewoude K., Dulhunty J., Blot S. Cuff pressure of endotracheal tubes after changes in body position in critically ill patients treated with mechanical ventilation // Am. J. Crit. Care. – 2014. – Vol. 23(1). – P. e1-8. https://doi.org/10.4037/ajcc2014489
- Danielis M., Benatti S., Celotti P., De Monte A., Trombini O. Il monitoraggio pressorio continuo della cuffia del tubo endotracheale: best practice in terapia intensive [Continuous monitoring of endotracheal tube cuff pressure: best practice in intensive care unit (in Italian)] // Assist. Inferm. Ric. – 2015. – Vol. 34(1). – P. 15-20. https://doi.org/10.1702/1812.19746
- Dauvergne J.E., Geffray A.L., Asehnoune K., Rozec B., Lakhal K. Automatic regulation of the endotracheal tube cuff pressure with a portable elastomeric device. A randomised controlled study // Anaesth. Crit. Care Pain Med. – 2020. – Vol. 39(3). – P. 435-441. https://doi.org/10.1016/j.accpm.2020.04.007
- Detsky M.E., Jivraj N., Adhikari N.K., Friedrich J.O., Pinto R., Simel D.L., Wijeysundera D.N., Scales D.C. Will This Patient Be Difficult to Intubate?: The Rational Clinical Examination Systematic Review // JAMA. – 2019. – Vol. 321(5). – P. 493-503. https://doi.org/10.1001/jama.2018.21413
- Kumar C.M., Seet E., Van Zundert T.C.R.V. Measuring endotracheal tube intracuff pressure: no room for complacency // J. Clin. Monit. Comput. – 2021. – Vol. 35(1). – P. 3-10. https://doi.org/10.1007/s10877-020-00501-2
- Coelho R. de M., de Paiva T.T., da Silva Telles Mathias L.A. In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes // Braz. J. Anesthesiol. – 2016. – Vol. 66(2). – P. 120-125. https://doi.org/10.1016/j.bjane.2014.06.012
- Duarte N.M.D.C, Caetano A.M.M., Arouca G.O., Ferrreira A.T., Figueiredo J.L. Insuflação de balonete de tubo traqueal por método subjetivo: desempenho de médicos residentes e especialistas em anestesiologia. Estudo prospectivo observacional [Subjective method for tracheal tube cuff inflation: performance of anesthesiology residents and staff anesthesiologists. Prospective observational study (in Spanish)] // Braz. J. Anesthesiol. – 2020. – Vol. 70(1). – P. 9-14. https://doi.org/10.1016/j.bjan.2019.09.010
- Li M., Yiu Y., Merrill T., Yildiz V., deSilva B., Matrka L. Risk factors for posttracheostomy tracheal stenosis // Otolaryngol. Head Neck Surg. –2018. – Vol. 159(4). – P. 698-704. https://doi.org/10.1177/0194599818794456
- Abubaker .J, Zia Ullah S., Ahmed S., Rehman Memon A.U., Abubaker Z.J., Ansari M.I., Karim M. Evaluating the Knowledge of Endotracheal Cuff Pressure Monitoring Among Critical Care Providers by Palpation of Pilot Balloon and By Endotracheal Tube Cuff Manometer // Cureus. – 2019. – Vol. 11(7). – Art. no. e5061. https://doi.org/10.7759/cureus.5061
- Marjanovic N., Boisson M., Asehnoune K., Foucrier A., Lasocki S., Ichai C., Leone M., Pottecher J., Lefrant J.Y., Falcon D., Veber B., Chabanne R., Drevet C.M., Pili-Floury S., Dahyot-Fizelier C., Kerforne T., Seguin S., de Keizer J., Frasca D., Guenezan J., Mimoz O.; AGATE Study Group. Continuous Pneumatic Regulation of Tracheal Cuff Pressure to Decrease Ventilator-associated Pneumonia in Trauma Patients Who Were Mechanically Ventilated: The AGATE Multicenter Randomized Controlled Study // Chest. – 2021. – Vol. 160(2). – P. 499-508. https://doi.org/10.1016/j.chest.2021.03.007
- Sevdi M.S., Demirgan S., Erkalp K., Akyol O., Ozcan F.G., Guneyli H.C., Tunali M.C., Selcan A. Continuous Endotracheal Tube Cuff Pressure Control Decreases Incidence of Ventilator-Associated Pneumonia in Patients with Traumatic Brain Injury // J. Invest. Surg. – 2022. – Vol. 35(3). – P. 525-530. https://doi.org/10.1080/08941939.2021.1881190
- Chenelle C.T., Oto J., Sulemanji D., Fisher D.F., Kacmarek R.M. Evaluation of an automated endotracheal tube cuff controller during simulated mechanical ventilation // Respir. Care. – 2015. – Vol. 60(2). – P. 183-190. https://doi.org/10.4187/respcare.03387
Ошибка: Контактная форма не найдена.